Subject: CMULTIS006-1 Segment: LowerLeg Modality: Compare MRI and CT Sphere Fitting - CT STL Marker Radius (mm) Coordinates [x,y,z] (mm) T1 9.42 75.8296, 221.0637, 1170.3869 T2 9.40 4.9650, 215.5120, 1134.4505 T3 9.51 -11.3693, 192.4694, 1155.1030 T4 9.56 29.9538, 216.1166, 971.9302 T5 9.46 32.4008, 216.2892, 848.9019 T6 9.53 3.9912, 187.3385, 826.4102 Distances between markers Marker 1 Marker 2 Distance (mm) T1 T2 79.65 T1 T3 93.03 T1 T4 203.75 T1 T5 324.44 T1 T6 353.01 T2 T3 34.99 T2 T4 164.43 T2 T5 286.86 T2 T6 309.33 T3 T4 189.26 T3 T5 310.23 T3 T6 329.09 T4 T5 123.05 T4 T6 150.59 T5 T6 46.38 Sphere Fitting - MRI STL Marker Radius (mm) Coordinates [x,y,z] (mm) T1 8.12 76.7495, 221.2512, 1171.8846 T2 7.60 4.5144, 215.4692, 1134.8597 T3 7.71 -11.5318, 192.4129, 1155.7029 T4 7.85 29.4225, 216.0165, 971.2448 T5 7.91 32.7356, 216.2965, 847.9684 T6 8.18 3.8807, 187.3430, 825.5223 Distances between markers Marker 1 Marker 2 Distance (mm) T1 T2 81.38 T1 T3 94.27 T1 T4 206.21 T1 T5 326.93 T1 T6 355.57 T2 T3 34.98 T2 T4 165.50 T2 T5 288.28 T2 T6 310.61 T3 T4 190.42 T3 T5 311.82 T3 T6 330.58 T4 T5 123.32 T4 T6 150.70 T5 T6 46.63 Differences between CT and MRI registration Marker 1 Marker 2 Distance Diff Angle(deg) T1 T2 -1.73 mm [-2.17 %] 0.26 T1 T3 -1.24 mm [-1.33 %] 0.43 T1 T4 -2.46 mm [-1.21 %] 0.26 T1 T5 -2.49 mm [-0.77 %] 0.05 T1 T6 -2.55 mm [-0.72 %] 0.08 T2 T3 0.01 mm [0.03 %] 0.57 T2 T4 -1.07 mm [-0.65 %] 0.09 T2 T5 -1.41 mm [-0.49 %] 0.13 T2 T6 -1.29 mm [-0.42 %] 0.07 T3 T4 -1.16 mm [-0.61 %] 0.20 T3 T5 -1.59 mm [-0.51 %] 0.05 T3 T6 -1.49 mm [-0.45 %] 0.01 T4 T5 -0.27 mm [-0.22 %] 0.40 T4 T6 -0.10 mm [-0.07 %] 0.18 T5 T6 -0.25 mm [-0.55 %] 0.45 Average 1.27 mm [0.68 %] 0.22 Maximum 2.55 mm [2.17 %] 0.57