Subject: CMULTIS006-1 Segment: UpperArm Modality: MRI Digitized Points Sphere Fitting Marker Radius (mm) Coordinates [x,y,z] (mm) H1 9.92 82.1227, 31.3823, 35.0247 H2 9.51 99.8393, 1.8759, -21.7929 H3 10.31 53.2641, 24.5894, 9.6983 H4 11.63 36.3421, -27.1020, -231.9576 Distances between markers Marker 1 Marker 2 Distance (mm) H1 H2 66.43 H1 H3 38.99 H1 H4 277.12 H2 H3 60.64 H2 H4 221.45 H3 H4 247.70 Sphere Fitting - MRI STL Marker Radius (mm) Coordinates [x,y,z] (mm) H1 8.42 82.3442, 31.5041, 36.5213 H2 7.95 100.4399, 1.9013, -21.3339 H3 8.27 52.9288, 24.9340, 10.0911 H4 8.28 35.8554, -27.5938, -234.3059 Distances between markers Marker 1 Marker 2 Distance (mm) H1 H2 67.46 H1 H3 40.09 H1 H4 281.07 H2 H3 61.44 H2 H4 224.50 H3 H4 250.56 Differences between digitized and MRI registration Marker 1 Marker 2 Distance Diff Angle(deg) H1 H2 1.03 mm [1.53 %] 0.35 H1 H3 1.10 mm [2.73 %] 0.89 H1 H4 3.95 mm [1.41 %] 0.05 H2 H3 0.81 mm [1.31 %] 0.54 H2 H4 3.04 mm [1.36 %] 0.07 H3 H4 2.86 mm [1.14 %] 0.06 Average 2.13 mm [1.58 %] 0.32 Maximum 3.95 mm [2.73 %] 0.89